Home » Application » Differential Pressure Transducer Calibration

Differential Pressure Transducer Calibration

The calibration for the differential pressure transducer mainly includes range adjustment and zero migration. To ensure the normal operation of the differential pressure transducer and the accuracy of differential pressure and flow measurement, users need to periodically check and calibrate it. The differential pressure transducer has a positive pressure terminal and a negative pressure terminal. In general, the pressure at the positive pressure terminal should be greater than the pressure at the negative pressure terminal so that the measurement can be performed normally. In the actual calibration process, users often mistake that the range of the differential pressure transducer can be changed by the HART communicator, and the zero point and the range adjustment can be performed without inputting the pressure source. In fact, this method only sets the range and does not meet the calibration purpose. The correct calibration should be to use the standard instrument for zero and range calibration in the case of external force, and carry out the necessary range migration according to the actual working conditions on site.
  • Preparation for differential pressure transducer calibration
The three valves manifold used with the differential pressure transducer is connected to the pressure guiding tube in the measurement. Under normal circumstances, the connectors of the pressure guiding tube and the differential pressure transducer should be disassembled, and then introduce the pressure source for calibration. However, this method is heavy in workload and the process of disassembling and installing the connectors may cause the pressure tube to break, resulting in a leak. Since all differential pressure transducers have an exhaust/drain valve or cock in the positive/negative pressure chamber, they can be calibrated without removing the pressure tube.
  1. Before calibration, close the positive/negative valves of the three valves manifold.
  2. Open the balancing valve, loosen the exhaust/drain valve or vent the cock.
  3. Replace the exhaust/drain valve or cock of the positive pressure chamber with a self-made joint.
  4. At this time, the negative pressure chamber keeps loose to pass through the atmosphere into it. The pressure source is connected to the self-made joint via a rubber tube.
  5. Close the balancing valve and check the air circuit seal.
  6. Connect the ammeter (voltmeter) and the handheld communicator to the output circuit of the differential pressure transducer.
  7. Start the calibration after powering up and warming up.
The adjustment status of the differential pressure transducer three valves manifold is shown below.
State of the three valves manifold in differential pressure transducer calibration
  • Zero calibration for the differential pressure transducer
1. Differential Pressure transducer zero calibration process

First, open the balancing valve of the differential pressure transducer. Then close the two shut-off valves. In this way, the differential pressure transducer can be calibrated. Take the ATO-PRES-2051 differential pressure transducer as an example to introduce the zero adjustment for the differential pressure transducer. Loosen the screws on the differential pressure transducer housing to expose the zero adjustment button. Note that there are two buttons, one for the zero adjustment button (ZE-RO) and the other for the default setting restoring button (SPAN). Pressurize the differential pressure transducer. When the pressure is applied to 4 mA, output the corresponding pressure value. Press the zero adjustment button 2s to check if the output is 4mA. The pressure transducer with a head will display ZEROPASS.

2. Range adjustment and zero migration for differential pressure transducer
  • Differential pressure sensor range adjustment
In practical applications, it is sometimes necessary to change the zero point or range of the differential pressure transducer due to changes in measurement requirements or measurement conditions. For this purpose, zero migration and range adjustment can be performed on the differential pressure transducer. The purpose of the range adjustment is to make the upper limit of the differential pressure transducer output signal correspond to the upper limit of the measurement range. The figure below shows the input/output characteristics before and after the range adjustment of the differential pressure transducer. The range adjustment is equivalent to changing the slope of the input/output characteristics of the differential pressure transducer. The adjustment from characteristic 1 to characteristic 2 increases the range; otherwise, the range decreases.
Differential pressure transducer input/output curves before and after upper limit adjustment
  • Differential pressure transducer zero migration
For differential pressure transducers, laboratory zero calibration is usually performed prior to commissioning. In actual measurement, in order to meet the on-site process requirements, it is often necessary to migrate the starting point of the measurement to a certain value (positive or negative). This is the zero migration. When there is no migration, the measurement starting point is zero. When the measured starting point changes from zero to a positive value, it is the positive migration. On the contrary, it is the negative migration. The purpose of zero adjustment and zero migration is to make the lower limit of the differential pressure transducer output signal correspond to the lower limit of the measurement signal. When Xin is zero, it is zero adjustment. When Xin is not zero, it is zero migration. The input/output characteristics of the differential pressure transducer before and after the zero migration are shown in the figure below. After zero migration, the input-output characteristics of the differential pressure transducer are shift a distance to the right or left along the coordinate axis. The slope does not change, that is the range of the differential pressure transducer does not change. If zero migration is used, combined with range compression, the measurement accuracy and sensitivity of the differential pressure transducer can be improved.
Input/output curves of differential pressure transducers before and after zero migration
Zero positive and negative migration refers to the adjustable range of the differential pressure transducer zero point, but it is different from the zero point adjustment. Zero adjustment is the adjustment when the input signal of the differential pressure transducer is zero and the output is not zero (lower limit). The positive and negative migration of the zero is adjusting the output to zero (lower limit) when the input of the differential pressure transducer is not zero. If the low pressure port of the differential pressure transducer has input pressure and the high pressure port does not, the adjustment when the output is adjusted to zero (lower limit) is called negative migration. If the high pressure port of the differential pressure transducer has input pressure and the low pressure inlet does not, the adjustment to adjust the output to zero (lower limit) is called positive migration. Migration is a zero adjustment when the differential pressure transducer has input. Therefore, the amount of migration is expressed as a percentage of its migration input signal or measurement range.
  • The work after the calibration of the differential pressure transducer is completed
After the zero and range calibration of the pressure transducer is completed, first drain the gas/liquid in the positive and negative pressure chambers. Then turn the exhaust/drain valve or plug back into place and wrap the raw tape to ensure that no leaks will occur.

I have an Opinion:

    
Plz Calculate (5 - 8) =
(Your comment will show after approved.)

You may also like: